RSD Heart and Thorax Phantom | Cardiac and Thoracic Imaging QA Tool
Short Description:
The RSD Heart and Thorax Phantom is an anthropomorphic phantom designed for evaluating myocardial infarct detectability, extent, and severity in nuclear medicine and PET imaging. It supports testing of partial volume effects, receptor quantification, scatter and attenuation correction schemes, and uptake change thresholds. Suitable for comparing 2D and 3D PET acquisition modes, it also supports image registration validation and mammo-scintigraphy technique assessment.
Key Features:
- Anatomically accurate heart model made from vacuum-formed shells
- Myocardial wall with two ports for independent flushing and filling
- Heart chamber volume of 284 ml and myocardial wall volume of 238 ml, without inserted defects
- Heart with hollow defects available for infarct simulation
- Thorax molded from polyurethane, tissue-equivalent, with a mass density of 1.10 g/cc
- Thoracic cavity volume of about 8,200 ml with heart, lungs, and liver inserted
- Removable breasts and breast tumors for mammo-scintigraphy studies
- Thorax overlay simulates an average male patient
- Base plate with O-ring seal and drain valve for easy draining
- Heart and lung subassembly can be disassembled for access and modification
Evaluation Capabilities:
- Partial volume effects
- Receptor quantification as a function of uptake ratios
- Scatter and attenuation correction schemes
- Thresholds for changes in uptake
- Comparison of acquisition modes, including 2D and 3D PET
Applications:
- Myocardial infarct detectability and severity assessment
- Cardiac PET and SPECT reconstruction optimization
- Partial volume and attenuation correction validation
- Protocol development for patient imaging
- Image registration testing and validation
- Mammo-scintigraphy technique evaluation
- SUV validation as a quantitative phantom
- Staff training and departmental imaging standardization
Manufactured by RSD. Supplied and serviced by Meditronix Corporation in India.